King BLB-环球(香港)科技有限开云
Product
Location:Home > Our Products > Manufacturers > Tannas & King > King BLB
Product information
King BLB
Category:Industrial Lubrication
Method Standard:ASTM D2983,IP 267, DIN51398
Application:Measures low-temperature viscosity of ATF’s, gear oils and hydraulic oils, and other fluid lubricants
Brand:Tannas-King
Description Features Test Methods Instrument Specifications

The King BLB measures low-temperature viscosity of ATF’s, gear oils and hydraulic oils, and other fluid lubricants using a patented SimAir® technique that allows the use of a small, less costly, constant temperature liquid bath. The only technique available that permits independent sample analyses eliminates the need for batch sample testing.

The BLB is a widely used low-temperature test for a variety of industry specifications.


· Unique Tannas SimAir® Test Cells provide proper sample temperature control for accurate, precise results and permits each test cell to operate independently, increasing productivity.

· Features a 12-position carousel with 12 independent sample cells

· BLB models: Capable of measuring a broad temperature range of +30°C to –70°C (BLB702) and +30°C to –40°C (BLB 701).

· SimAir®/DIN model: Capable of measuring a broad temperature range of +30°C to –55°C.

· Fully self-contained internal hermetic refrigeration system – no external refrigeration units or controllers needed.

· Digital temperature controller keeps bath temperature within +/- 0.1°C.

· Small benchtop footprint with large illuminated viewing window.

· Includes top mount for viscometer (viscometer available separately).

· The internal air system provides a continuous blanket of dry air over samples to eliminate moisture buildup.

· Safety features include a high-pressure cutout, high-temperature limits, and low-level shutdown.

· CE Mark


· ASTM D2983

· ASTM D97

· ASTM D2500

· ASTM D5853

· IP 267

· DIN 51398


Dimensions

(W x D x H)

Benchtop: 42 x 53 x 66 cm

(16.5 x 21 x 26 inches)

Weight

~68 kg | (150 lbs.)

Voltage

BLB 701: 220 VAC, Fused: 15 amp.

BLB 702: 220 VAC, Fused: (2) 15 amp.

BLB-DIN: 220 VAC, 15 amp.

Frequency

50 or 60 Hz., Single-Phase

Cooling Capacity

600 Watts at 0°C | 300 Watts at –40°C | 100 Watts at –70°C

Heating Capacity

600 Watts

Temperature

Range

BLB 701: +30°C to –40°C (86°F to –40°F)

BLB 702: +30°C to –70°C (86°F to –94°F)

BLB-DIN: +30°C to –55°C (86°F to –67°F)

Temperature Control

±0.1°C | Digital readout to 0.01°C for BLB-DIN

Bath Cooling Rate

BLB 701: 20°C/ hour average

BLB 702: 30°C/ hour average

BLB-DIN: 60°C/ hour average

Refrigerant

R410A (BLB701) | R507 HFC / R508B (BLB702 & DIN)

Compressor

1/3 horsepower

Viewing Window

~41 x 23 cm | 16 x 9 inches

Testing Capacity

12 sample carousel (ASTM 2983) 4 sample carousel (ASTM D97 + D2500)

Bath Size

~3.8 liters (1 gallon)

Bath Design

Methanol or Ethanol within Dewar flask

Cabinet Material

Powder Coated Aluminum

Safety

High pressure cutout | High temperature limit

Low Level Indicator | Very Low Level Safety Shutdown

CE Mark

Shipping Weight

~122 - 136 kg | (270 - 300 lbs.)

Shipping Dimensions

(W x D x H)

81 x 76 x 104 cm

(32 x 30 x 41 inches)